ChemStarMSQuickGuide
1.ANALYSISUSINGMASSSPECTROMETER
QuickStartupGuideforThermostarPfeiffer-MS
RemovethetwoscrewsandthelockingscrewlocatedatthefrontbottomoftheThermostarMSbeforeoperatingtheunit.
1.Makesurethatthereissufficientlengthofcapillaryextendingoutofthecapillarytubing.Atleast2inchesofthecapillaryisrequiredtoextendout.
2.Connectthecapillarytotheback(analyticalvent)oftheChemStar.Refertofigure1,2and3fortheconnection.Figure2indicatesthedirectionofthegraphiteferrule.
3.OncetheMS-capillaryconnectionstotheChemStarhasbeenestablishedturnonthepowertotheMSbyuseoftheswitchonthebackoftheThermoStar.Thefanwillturnonatthispoint.
4.OnthefrontdisplayscreenoftheThermoStar,(figure4)presstheOK/SELbuttonthreetimes(youwillhearabeepsound)toturnontheturbopump.Theturbowillpumpdownuntiltherotationspeedreaches1500Hzor90000rpm.Thatwouldbeabout2-3minutestoreachfullspeed.
5.Thenormalpressurewouldbearound1.28E-05mbaruponfullspeedoftheturbopump.
6.Iftheinstrumentisstartingupforthefirsttime,thenabake-outoftheunitisnecessary.
7.Toperformthebake-outmanually,thefollowingstepsareneeded.
a)Openthe“Quadera”softwarefromthecomputer.
b)Seefigure5oncetheQuaderaopensup.
c)Clickon“PTMxxxxx”asshowninfigure6toopenupthelivemode.
d)Thenselectthe“SEMMID”asshowninfigures7and8.
e)Thescreenshowingmultipleiondetectionasshowninfigure9willopenup.
f)ClickonthearrowtoturnONthefilament.(AmessagewillaskifyouwanttoturnOnthefilament).Clickyestoturnonthefilament.
g)ThefilamentmustbeturnedONwhenperformingabake-out.
h)OnthefrontdisplayscreenoftheThermoStar,selecttheOK/SELbuttontomovebetweenfunctions.Onthefunctionstab,selectthedownarrowkeytoreachbake-out.Selectthebake-outmanual,andthenhitoktostartthemanualbake-outsession,preferablyovernight.
i)Tostopthebake-out,clickonOK/SEL,thenmovebetweenfunctionstogotobake-outmanual>bake-outstop.Thiswillstopthebake-outsession.
8.Now,youarereadytoscantheMSsignals.Toeditthemasses,clickon“edit”tab.(seefigure9).Thiswillthenopenupthemassesscreenasshowninfigure10.
9.Thesoftwareispre-loadedwithupto7masses.Toaddtheeighthmass,performthefollowing:
a)Selectthe“MID”scanasshowninfigure11.
b)Thenclickon“AddMIDtask”,andthendraganddropitontothemeasurescreen.
c)Theaddedtaskwillbedisplayedonthemasseslistonthemeasurescreen.Seefigure12.
d)Nowgotothe“edit”tabandselecttheeighthmassandchangethe“mass”and“display”totherequiredmassnumberandnamethatisneeded.
10.Afterenteringthemasses,gotothemeasuretabandclickthemeasurearrowbuttontostartmeasuring.Andthiswillpullupthescreenasshowninfigure13.
11UsetheChemStarsoftwaretoflowgasesthroughtheTCDandtheanalyticalventtodetectthemasses.
12Openthe“Shamus”softwarebyclickingonthe“Shamus”icononthedesktopofthecomputer.Thiswillopenupthescreenasshowninfigure14.
13TheshamusisthesoftwarecommunicationlinkbetweenQuaderaandtheChemStarsoftwareviaOPC.Themassesareautomaticallylistedontheshamusscreen.Toupdateanymasschangesonthe“Quadera”totheshamussoftware,clickonthe“getparameter”button.
14OntheChemStarsoftwarescreen.Goto“accessories”andthenselect“MassSpectrometer”.Thiswillopenupthescreenshowninfigure15.ClickonthelinkONbuttontoestablishcommunicationoftheQuaderaandtheChemStarsoftware.
15WhenanexperimentisinitializedwiththeChemStarsoftware,theMSsignalswillbeautomaticallyrecordedtotheexperimentrundata.
SHUTDOWN
Toensurethatapropershutdownisdone,followthestepsbelow:
Closetheshamussoftwarelinkscreen.ThenturnoffthemeasurementandthenthefilamentshouldbeturnedoffandtheSEMdetectormustbeturnedoff.Refertofigure9.ThenclosetheQuaderasoftware.
16OnthefrontdisplayscreenoftheThermoStar,presstheOK/SELbuttontoselectFunctions>ShutDown.Awarningsignwillsay“TurningOffvacuumwithsensorON,mayharmfilament”.ThenpressOKtoshutdown.OncetheOK/SELkeyisselected,theturbopumpwillshutdown.Thismaytakeupto10minutes.Oncetheturbopumpisshutdownthefrontpaneldisplaywillsay“systemok”.
Figure1
Figure2
Figure3
Figure4
Figure5
Figure6
Figure7
Figure8
Figure9
Figure10
Figure11
Figure12
Figure13
Figure14
Figure15