Purcon样品预处理器
PURCON
PURCON
样品预处理仪
1
PURCON
目录
1.介绍.......................................................................................................................................................................................3
1.1概要...............................................................................................................................................................................3
1.2操作...............................................................................................................................................................................3
2.安全.......................................................................................................................................................................................6
2.1授权使用.......................................................................................................................................................................6
2.2一般安全规范...............................................................................................................................................................6
2.3警示说明和符号..........................................................................................................................................................6
3.调试.......................................................................................................................................................................................8
3.1安装PURCON.............................................................................................................................................................8
3.2连接进料口...................................................................................................................................................................9
3.3连接滞留管口...............................................................................................................................................................9
3.4连接排泄管口...............................................................................................................................................................9
3.5安装PURCON空气排放口(选配).......................................................................................................................9
3.6设置手动阀.................................................................................................................................................................10
3.7电源.............................................................................................................................................................................10
3.8连接分析仪.................................................................................................................................................................11
3.9插入泵导管.................................................................................................................................................................11
3.10开始样品预处理......................................................................................................................................................11
4.操作.....................................................................................................................................................................................12
5.保养.....................................................................................................................................................................................12
5.1移位泵导管.................................................................................................................................................................12
5.2更换泵导管.................................................................................................................................................................13
5.3清洗PURCON..............................................................................................................................................................13
5.3.1外壳...................................................................................................................................................................13
5.3.2样品预处理......................................................................................................................................................13
5.4清洁空气过滤器........................................................................................................................................................15
5.5更换保险丝.................................................................................................................................................................16
5.6仪器暂停使用.............................................................................................................................................................16
6.故障索引............................................................................................................................................................................16
7.技术参数............................................................................................................................................................................16
8.备用部件和附件................................................................................................................................................................18
9.清洁剂安全数据单............................................................................................................................................................18
2
PURCON
1.介绍
1.1概要
PURCON样品预处理仪对分析仪的样品成分分析进行预处理。
PURCON采用最现代化的技术使得操作可
以连续进行而几乎不需保养。
与传统的样品预处理仪相比,PURCON的保养几乎可以忽略。
由于样品在PURCON中的滞留时间很短,
所发生的预处理对样品没有产生任何影响。
样品预处理连续进行,处理过的渗透液不含细菌和固体使分析
仪高度可靠。
采用PURCON进行样品预处理后,分析仪所需的保养也大为减少。
图1 PURCON样品预处理仪
1.2操作
图2操作功能元件
3
PURCON
进料
进料通过一软管连接至进料管口(PURCON外壳上)。
进料能通过泵抽取或把样品容器放置在比PURCON
高的位置上通过重力流入PURCON。
外壳
样品预处理在PURCON壳内进行。
滞留物
任何过量样品(滞留物)将通过废置管口排出。
当化学品不再添加时,任何过量样品将被废置。
渗析
渗析泵通过软管为分析仪和PURCON传送渗析液,其连接至渗析管口。
传送量能通过4档选择开关控制,因此能最佳地满足需求。
清洗
在使用6-9个月后,样品预处理仪必须清洗。
从填充管口注入清洁剂并通过排放口废置。
在污水处理厂内
部所排放的清洁液能够进行自我处理。
然而,在预处理饮用水和地表水时,清洁液必须单独处置。
仪器内部
在仪器内部进行样品预处理。
操作所需气体由压缩机提供。
由阀控制的4个渗析道交替使用。
PURCON
由渗析管口连至分析仪。
阀门设置
两个手动阀门控制进料流的开关或抽空样品预处理。
图3正常操作的阀门设置
图4抽空样品预处理仪阀门设置
4
PURCON
图5清洗样品预处理仪阀门设置
5
PURCON
2.安全
该操作手册介绍用户在仪器调试、操作和保养时要遵守的基本事项。
用户在使用仪器前必须详细阅读该操
作手册。
把操作手册放在仪器附近以备随时查阅!
2.1授权使用
该仪器仅限使用:
对要使用分析仪进行成分分析的样品进行预处理。
必须遵守技术参数一节中的技术规格。
其它任何使用均为非授权使用。
2.2一般安全规范
该仪器制造和检验符合电子测试仪的相关安全法规。
分析仪出厂时处于良好的状态。
功能和操作的安全性
在使用仪器时,只有认真遵守操作手册中的安全规范,仪器的功能和操作仪器的安全性才能得到保障。
只有满足技术参数所规定的气候条件,仪器的功能和操作安全性才有保障。
如果仪器从低温移向高温,冷凝现象可能会导致仪器故障,这时要等仪器温度达到室温后才开始测试。
Warning:
仪器的打开、校准、保养和维修工作,非WTW授权技术人员不得执行。
EN61010-1
该仪器符合EN61010-1所有的要求,保护等级I。
欲了解该类保护的更多信息,请参阅技术参数。
所有超出操作手册中描述的举措,特别是安装工作,必须由训练有素的人员执行。
更换保险丝也必须遵守该守则。
安全操作
若仪器不能安全操作,则停止工作以避免错误操作!
出现下列情况不能安全操作,如:
●
●
●
●
仪器运输过程中受损
仪器保存不当
仪器受损,目视可见
仪器操作有误
若尚存疑问,请联系供应商。
符号说明
Warning:
指示用户必须阅读—为了用户及他人的安全亦为了保护仪器免受损害。
Caution:
指示小心危险电压。
Note1:
指示用户必须注意的事项。
2.3警示说明和符号
危险物质规则中的危险符号:
强腐蚀性。
谨防小孩触碰该物质。
操作时戴好合适的护手套和护目镜/面罩。
如物质不慎入眼,请立即用水冲洗并咨询医生。
危险物质规则中的危险符号:
该标识为危险指示符号。
其一般应用在相关化学品容器上或相应的操作手册中。
紧跟标识的指示必须严格遵守。
命令符号,戴上护目镜。
该标识表示用户必须戴好护目镜。
该符号在说明书中警示对人体健康,特别是眼睛的危险操作。
PURCON
命令符号,戴好护手套。
该符号指示用户必须戴好护手套。
该符号在说明书中警示对人体健康,特别是手的危险操
作。
命令符号,穿上防护衣。
该符号指示用户必须穿上防护衣。
该符号在说明书中警示对人体健康有危害的危险操作。
7
PURCON
3.调试
货品清单
●
●
PURCON样品预处理仪
操作手册
Warning:
PURCON仅能由经过良好训练的人员安装。
钥匙必须安全保管且仅能由资格电工使用。
图6尺寸
3.1安装PURCON
1
2
取4个.螺丝钉(8mm)以备仪器安装。
用4个.螺丝钉把固定片固定在墙上或PURCON竖架(附件)。
Warning:
确保.螺丝钉强度足够。
若PURCON安装到墙上,确保墙的强度足够支撑PURCON的重量(见
技术参数一章)。
Warning:
确保PURCON安装在分析仪的附近以避免进料路线过长延长迟滞时间。
Warning:
凳面必须与地面固定。
未经固定不能使用,固定采用所提供的螺丝钉。
PURCON
图7PURCON及凳面
3.2连接进料口
用3/4”软管(ID19mm)连接进料传送器到PURCON进料管口。
Warning:
设置进料传送器传送量请遵循操作说明书技术参数一节。
3.3连接滞留管口
1
2
取50mm的排泄管。
连接排泄管至PURCON的滞留管口。
排泄系统必须畅通。
PURCON不消弱进料。
Warning:
使用排泄管时采用生胶带,确保生胶带粘贴牢固。
若使用软管,需用管夹夹紧。
3.4连接排泄管口
排泄管口用来抽空样品预处理仪。
清洗期间,排出清洁液。
正确处置清洁液。
Warning:
清洁液不能排入饮用水或地表水中。
如安全数据表所描述处置清洁液。
3.5安装PURCON空气排放口(选配)
在排放管不完全畅通的情况下我们建议用户安装PURCON气体排放口。
采用气体排放方式可以避免形成
过压以至阻碍滞留物的排放。
PURCON气体排放口安装在填充管上替代填充管帽。
图8PURCON空气排放口
9
PURCON
安装
1
2
3
4
5
6
暂停进料传送器。
关闭渗析泵(把旋转开关打至0档)。
关闭PURCON电源。
旋开机器外壳顶部的填充管帽。
如图6安上PURCON空气排放口并用手旋紧。
消除异味,把合适的DN30导管插入喷嘴管口且让另一端敞口或排入排放系统。
3.6设置手动阀
PurCon外壳上的阀具有多种操作模式。
Warning:
调试仪器时确保设置正确(见图9)。
图9进料口打开,出口关闭
3.7电源
Caution:
连接仪器前请电工检查交流电插座的负荷能力。
插座必须接地。
避免连接分线。
1
2
取出随机电线。
连接PurCon交流电源。
Caution:
PurCon铭牌上的数据(标明线电压)必须与交流电插座电压一致。
详见技术参数一章查阅连接
数据。
图10PurCon交流电插座(外壳俯视)
10
PURCON
3.8连接分析仪
1
2
取出随机的渗析软管。
连接分析仪至仪器前面的渗析管口。
Warning:
从PurCon至第一个分析仪的渗析软管长度不能超过2米。
3.9插入泵导管
按如下步骤进行:
1
2
3
4
5
按仪器夹钳从固定器上取下泵盒。
用泵导管连接泵盒中的第一和第二个制动器。
放泵盒回固定器并卡住。
把泵盒安装在螺钉杠上。
重复该过程安装其它4个渗析管。
图11插入泵导管
3.10开始样品预处理
PurCon已准备待测。
在接通电源之前,请检查如下几点:
1
2
3
4
5
所有的导管是否夹紧?
滞留物排放管口是否密封,连接是否正确且足够紧?
手动阀是否为操作样品预处理仪而正确设置?
所有渗析泵螺钉杠是否拧紧?
渗析管是否正确连接至分析仪?
按如下步骤启动样品预处理仪:
1
2
3
打开交流电开关(绿色电源灯亮)。
打开进料泵(绿色控制灯亮)。
打开渗析泵。
PurCon开始样品预处理。
首先,系统开始通风。
在启动渗析泵约8分钟后渗析传送开始,然后可以启动
分析仪。
Warning:
调节渗析量以满足用户需求。
如下设置可选:
位置1
位置2
位置3
位置4
操作一个分析仪
操作二个分析仪
操作三个分析仪
仅用来加速样品预处理仪的通风过程。
11
PURCON
4.操作
PurCon操作几乎不需保养。
在正常操作情况下,仪器仅需有限的保养。
详情请参阅保养一章。
5.保养
与传统的样品预处理仪相比PurCon所需保养很少。
它们仅限于如下保养:
●
●
●
●
移位泵导管
清洁压缩机的空气过滤器
更换泵导管
清洁PurCon
如下表格列出保养概况,依据样品质量,保养间隔可改变。
措施
月
季
半年
移位泵导管
清洁空气过滤器
更换泵导管
清洁PurCon
√
√
√
√
5.1移位泵导管
泵导管必须移位以避免过重的机械负荷。
若定期进行移位其使用寿命可延长至6个月。
在更换导管前可进
行5次移位。
按如下步骤进行泵导管移位: